Used cryopumps sold by Capovani Brothers are tested at our Scotia, New York facility before shipment. Most inventory sold as working and warranted is tested to the procedures on these pages. Contact our sales team for details.

What We Check

Our Cryopump Testing Process

  • Overall assessment performed; cryopump made operational in current condition to identify problematic areas before testing begins
  • Main sealing surface visually inspected for scratches, dings, and wear. Conflat knife edges must be in near-perfect condition; nicks, dings, or deep scratches are not tolerated
  • 15K and 80K arrays inspected for corrosion and degradation; replaced as necessary
  • Radiation shield inspected for corrosion and degradation; repaired or replaced as necessary
  • Pump can inspected for corrosion and deformation. Deformed pump cans are not sold; cans needing electro-polishing are sent out
  • Heat pump inspected for wear and operated to assess performance and audible noises indicating mechanical wear; rebuilt (interior cleaned, seals and bearings replaced) if needed
  • Regeneration purge fitting inspected, verified, and seals replaced as necessary
  • Pressure relief valve inspected, verified, and seals replaced as necessary
  • Hydrogen vapor bulb assembly inspected, verified, and replaced as necessary
  • Silicon diode temperature sensor inspected, verified, and replaced as necessary
  • Cryogenic performance verified and leak checked